2) Check the integrity of the supply cable
While checking, focus namely of the cable terminations, in place where the cable
enters the termination. This is the place where the external insulation layer often crackles.
Verify that the insulation is compact throughout the cable’s length and that the cable shows no
signs of breaking or notching. Any occurrence of bulging on the cable is also unallowable, as
it demonstrates damage of the conductors inside. Record the findings in the PSTC protocol by
means of checking the cell “Satisfactory”.
3) Check the functioning of the signalizing and controlling elements
Examine all the optical signalizing elements for missing function descriptions. Check
the mechanical running of the controlling elements (the buttons must be easy to push, rotation
switches must exert adequate resistance in switching and they must not allow for turning the
entire revolution). In membrane buttons, check the integrity of the top foil with function
descriptions, as it must not be perforated in any place. The LCD display must not show signs
of mechanical damage. After switching the apparatus on (in accordance with the operation
manual), check whether all the optical signalizing elements meet their function. Record the
findings in the PSTC protocol by means of checking the cell “Satisfactory”.
4) A - Measure the value of the leakage current of the apparatus and the resistance of the
protective grounding (for ARCOMED Siramed
µSP6000 and Volumed µVP7000)
To measure the protective conductor’s resistance, proceed in accordance with the user
manual of REVEX 2051. Detailed measurement procedure can be found in chapter 4.1.1. on
page 13. During the measurement, follow the instructions of the trainer and always proceed
only in accordance with the manual of REVEX 2051. Record the result into the PSTC
protocol and compare with the recommended maximum value stated in the protocol. Confirm
whether the machine passed.
To measure the value of the apparatus’ leakage current, proceed in accordance with
the user manual of REVEX 2051. Detailed measurement procedure can be found in chapter
4.4.1 on page 26. During the measurement, follow the instructions of the trainer and always
proceed only in accordance with the manual of REVEX 2051. Record the result into the
PSTC protocol and compare with the recommended maximum value stated in the protocol.
Confirm whether the machine passed.
4)
B - Measure the value of the insulation resistance of the apparatus and the leakage current
via the cover of the apparatus