User manual
(for SEV LITOVEL 2P and POLYMED ID 20/50)
To measure the apparatus’ insulation resistance, proceed in accordance with the user
manual of REVEX 2051. Detailed measurement procedure can be found in chapter 4.3.1 on
page 22 (for SEV LITOVEL 2P) and 4.3.2 on page 23 (for POLYMED ID 20/50). During the
measurement, follow the instructions of the trainer and always proceed only in accordance
with the manual of REVEX 2051. Record the result into the PSTC protocol and compare with
the recommended maximum value. Confirm whether the machine passed.
To measure the leakage current via the apparatus’ cover, proceed in accordance with
the user manual of REVEX 2051. Detailed measurement procedure can be found in chapter
4.4.1 on page 26 (for SEV LITOVEL 2P) and 4.4.2 on page 27 (for POLYMED ID 20/50).
During the measurement, follow the instructions of the trainer and always proceed only in
accordance with the manual of REVEX 2051. Record the result into the PSTC protocol and
compare with the recommended maximum value. Confirm whether the machine passed.
5) Carry out the verification of the apparatus’ functional parameters, according to the items of
the particular PSTC protocol
For a complete analysis and testing of the infusion technology, use the apparatus
SECULIFE IF. Prior to the actual measurement of the infusion technology parameters, set up
the measurement apparatus according to the operation manual of SECULIFE IF, page 11,
chapter 3. Two modes will be used for the measurement:
The mode for measuring occlusal pressure is described in chapter 4.3.2 on page 27.
Before you start measuring, make sure the system is correctly irrigated and that the three-port
valve is in the “ON” position, see chapter 3.2, page 13 of the operation manual. Record the
result into the PSTC protocol and compare with the recommended maximum value. Confirm
whether the machine passed.
The mode for measuring the flow is described in chapter 4.3.1 on page 27. Before you
start measuring, make sure the system is correctly irrigated and that the three-port valve is in
the “ON” position, see chapter 3.2, page 13 of the operation manual. Record the result into the
PSTC protocol and compare with the recommended maximum value. Confirm whether the
machine passed.
The actual set up of the measurement mode can be found in chapter 4.4 of the
operation manual. Occlusal pressure measurement can be launched to advantage during the
flow measurement.
65