To check the bubble detector, turn the infusion bottle in such a way as to allow for a
small amount of air (a bubble) to be sucked into the infusion pump’s infusion set. Then wait
whether the alarm sets off when the bubble passes through the infusion pump.
Set up the infusion pumps and linear infusion pumps according to the measured
parameters in the particular PSTC protocol. The actual controlling and set up for the
individual parameters can be found in the respective operation manuals for each apparatus.
Measuredresults
Record the results in the individual protocols. Be particular about the level of the
records, as in reality, crossed out PSTC protocol is legislatively non-permissible.
Checkupquestionsregardingthegivenissue
What is the occlusal pressure?
What is the functional difference between the infusion pump and the linear infusion
pump?
What is the sense of PSTC?
Which legislative document describes the necessity of carrying out the PSTC?
Referencestousedandrecommendedinformationsources
[5.1] Wesbter, J.G. ed. Encyclopedia of Medical Devices and Instrumentation. Wiley.
[online]. c1999-2009, poslední aktualizace 17. 8. 2008 [cit. 2009-05-12]. Dostupné z
WWW: http://mrw.interscience.wiley.com/emrw/9780471732877/home/
[5.2] Rozman, J. a kol. Elektronické přístroje v lékařství. Praha: Academia, 2006. 406 p.
(in Czech)
[5.3] Zákon č. 123/2000 Sb., zdroj: SBÍRKA ZÁKONŮ ročník 2000, částka 36, ze dne
12.5.2000. (in Czech)
[5.4] ČSN EN 60601-1-2. Zdravotnické elektrické přístroje: Všeobecné požadavky na
bezpečnost. 02/2003. (in Czech)
[5.5] ČSN EN 62353. Zdravotnické elektrické přístroje: Opakované zkoušky a zkoušky po
opravách zdravotnických elektrických přístrojů. 8/2008. (in Czech)
66