application of the artificial lung ventilation, together with other risks, such as a greater
infection risk, oxygen toxicity in its increased faction in inspired gas, etc. Such lung distress is
identified as the “lung distress caused by artificial lung ventilation”. The modern trend thus
prefers the so called protective ventilation modes, the main goal of which is to reduce the
adverse effects of artificial lung ventilation on the patient.
Thecontentsandtheaimofthemeasurements
The aim of this task is to introduce the artificial lung ventilation to the students.
Particular options will be demonstrated on the Veolar apparatus manufactured by Hamilton
Medical (Fig. 6.1), which will be connected to the breathing simulator. Veolar represents the
conventional artificial lung ventilation. Within the framework of the measurement, you will
become acquainted with the controlling and monitoring panels, and also with the alarms
guarding the basic parameters used during patient ventilation.
Fig. 6.1: Conventional ventilator VEOLAR manufactured by Hamilton Medical.
Ventilationmodes
After switching the apparatus on, it is necessary to select the ventilation mode. The
mode is selected by holding the button with the name of the ventilation mode in the left part