User manual
Spontaneous breathing ventilation mode (SPONT)
This mode is designed for patients with spontaneous breathing activity and is used for
its support and greater effectiveness. VEOLAR offers the following support options: supply
of air enriched with oxygen, continuous positive airway pressure (CPAP) and inspiratory
assistance during inspirium (pressure support). In this working mode, the ventilator functions
as a “DEMAND” system, providing the patient with the needed gas flow in selected pressure
values of CPAP or inspiratory assistance. The flow is controlled in such a way as to ensure
expirium of sufficient length. In this working mode, the patient determines the required
breathing volume and the breathing frequency, having given values of the inspiratory
assistance’s positive pressure. It is possible to select “back up” ventilation in case of apnoe.
In the SPONT mode, the following parameters may be set up, marked by luminous
spots: Pressure Trigger - pressure trigger, CPAP - continuous positive airway pressure,
Oxygen - oxygen fraction in the inspiratory mix, Flow Trigger - flow trigger.
MMV ventilation mode
Minimum Minute Ventilation. This ventilation mode allows for spontaneous breathing
with automatic inspiratory assistance, the pressure value of which is selected directly by the
ventilator. In MMV, the patient gets, if necessary, breathing assistance on the level of
inspiratory assistance, yet no controlled breaths. MMV ensures the patient’s minute
ventilation automatically on the level of the pre-set minute ventilation.
In this mode, the following parameters may be set up, marked by luminous spots:
Pressure Trigger - pressure trigger, CPAP - continuous positive airway pressure, Oxygen -
oxygen fraction in the inspiratory mix, Flow Trigger - flow trigger, MMV - minimum minute
ventilation - to be set up via the
∧∨ buttons, followed by pushing the YES button on the
Alarm panel. Just like in the SIMV or Spont modes, the trigger must be set optimally. If the
trigger is switched off, the apparatus will automatically set the alarm off.
During the first breaths, the patient may breathe in the way corresponding to the set
inspiratory assistance. The ventilator compares the volume of the patient’s first eight breaths
by converting them to one minute with the set-up MMV value. If there is a deviation of the
minute ventilation from the set-up value, the balancing of the deficit starts by means of
increasing the pressure support. A change of the pressure support is carried out gradually, step
by step, and depends on the size of the difference between the actual and the required minute
ventilation.
70