Thecontentsandtheaimofthemeasurements
High-frequency electrosurgery provides means for fine and accurate surgical
interventions on vascular tissues. Using this method prevents undesirable extensive damage to
the surrounding tissue. To achieve the best possible results, it is necessary to work with the
unit’s performance set-up as gently as possible. The task demonstrates using ESU with the
help of a specialized tester, the operation of which is based on measuring the working current
and supplied delivery. The aim of this task, therefore, will be to learn about and verify the
principle of the apparatus’ operation, to measure its functions by means of the tester, to find
out operation characteristics and current dependencies.
Measurementtasks
Prior to the actual measurement, become acquainted with the operation and the parts of the
electrosurgical apparatus (SMT BM CLINIC 170W, [5]) and of the analyzer of electrosurgical
devices (RF 303, [6]). Suggest and consult with the trainer all the measurement schemes and
connection of the individual apparatus. Work with the operation manuals [5, 6] available on
the subject’s website.
1)
Measure the output power in dependence on the size of the load resistance by means
of the tester (RF 303) for monopolar and bipolar modes and for all the four functions.
2)
Measure the passing current during section and coagulation by means of the tester (RF
303), for monopolar and bipolar modes at different ohmic load sizes.
3)
Use the tester (RF 303) to measure the leakage currents by the applied part.
4)
Elaborate a protocolar record on the carried out measurements and the safety-technical
check of the electrosurgical apparatus.
Measurementprocedure
To accomplish all the aims of this task, it is necessary to first become acquainted with
the operation of the electrosurgical generator, the electrosurgical apparatus’ tester, and the
electrical safety tester. First, learn how to connect the necessary accessories to the apparatus,
such as the foot switch, the neutral electrode and the active electrode applicator. (See Fig. 7.1
and 7.2). Become acquainted with the activation principle of this electrosurgical unit [7.5].
Further on, study the principle of the RF 303 tester operation, the means for connecting ESU
84