Fig. 7.2: Rear panel of the electrosurgical apparatus CLINIC 170 W [7.5].
Legend:
1 - pneumatic trigger connector - connecting the foot switch
2 - power cord plug
3 - power switch
4 - manufacturer’s label
Outputpowermeasurementindependenceonthesizeoftheloadresistancefor
monopolarandbipolarmodesandforallthefourfunctions
Connect the neutral electrode into the main panel of the electrosurgical unit. Connect
the interconnecting cable to it by means of a crocodile clip, and plug its second end into the
left blue socket on the tester. Plug the interconnecting cable into the active electrode’s socket,
and plug its second end to the yellow socket on the tester. Switch the apparatus into the
monopolar mode (see Fig. 7.1). Connect the foot switch (see Fig. 7.2). Use the mode switch
No. 10 (Fig. 7.1) to set up the functions of section, mixed section, coagulation and
microcoagulation subsequently. Use the intensity regulator No. 5 (Fig. 7.1) to set the output
performance to MAX. Set up the tester gradually for each function to the resistance of 50 Ω,
100 Ω, 200 Ω, 500 Ω, 750 Ω. Resistance is set on the tester via the OHMS SELECT + or -
button and by selecting the respective value, which is featured on the display (Fig. 7.3). Make
sure you have really selected the performance measurement. If not, use the ENT button to
select W (see Fig. 7.3).
86