function, namely with leakage currents and functional currents flowing from the high-
frequency apparatus via undesirable paths through the patient or the attending staff.
When using an electrosurgical apparatus, direct connection of the patient with the
apparatus cannot be avoided. As the ESU is powered by the
230 V/50 Hz electrical
distribution network, there is a danger in the form of leakage currents on the network
frequency (and their higher harmonic components). Leakage currents are understood as
parasite currents flowing between mutually insulated parts of the apparatus, mostly through
parasite capacity couplings between these parts.
Connect the active electrode of the electrosurgical apparatus in the right grey socket,
and the neutral electrode in the left blue socket. Then connect the yellow and the left green
socket, and the right blue and the left grey socket via a short connecting cable. (Note: yellow
socket = active, green = ground, blue = dispersion, grey = leakage currents). Set the resistance
on the tester to 200 . After that, make sure you really selected the current measurement. If
not, then use the ENT button to select mA (Fig. 7.3). The electrosurgical unit is switched to
the monopolar mode. If everything is right, activate the apparatus via the foot switch and
measure the ESU leakage currents in all the four modes gradually (section, mixes section,
coagulation, microcoagulation). Use the middle knob on the ESU main panel to gradually
select the performance from the MIN position all the way to the MAX position (Fig. 7.1 and
7.2).
Proceed likewise for the bipolar mode measurements, only following the instructions
in the operation manual [7.5] relating to operating the apparatus in bipolar mode.
For all the measurements, make use of the possibility of connecting the RF 303 tester
to the oscilloscope and display the course of the actuating signal in the individual modes and
functions of the electrosurgical generator CLINIC 170W. Record the measured results in Tab.
7.7 and 7.8.
Apparatusandaidsused
Electrosurgical generator SMT BM CLINIC 170W, neutral electrode for repetitive
use, foot switch and interconnecting conductors (Speciální Medicínská Technologie,
s.r.o., Czech Rep.)
Analyzer of electrosurgical devices RF 303 (Fluke Biomedical, USA)
Interconnecting conductors and crocodile clips (RF 303 accessories)
89