Non-Invasive Blood Pressure Simulator NIBP-1010 NIBP-1020 USER MANUAL
BC BIOMEDICAL NIBP-1000 SERIES TABLE OF CONTENTS DESCRIPTION............................................................................................................. 3 OVERVIEW.................................................................................................................. 5 KEYS............................................................................................................................ 7 SCREENS.......................................................................
WARNING All connections to patients must be removed before connecting the Device Under Test (DUT) to the Simulator. A serious hazard may occur if the patient is connected when testing with the Simulator. Do not connect any leads from the patient directly to the Simulator or DUT. NOTICE BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT.
BC GROUP NIBP-1000 SERIES NON-INVASIVE BLOOD PRESSURE SIMULATORS The Model NIBP-1000 Series is a family of Microprocessor based, High Precision NonInvasive Blood Pressure (NIBP) Simulators. The units are small, easy to use and have multiple features to fit many different applications. The NIBP-1020, offers ECG waveforms that are full QRS and respiration waveforms that look real.
AVAILABLE MODELS: NIBP1010 NIBP1010-P NIBP1010-BP NIBP1020 NIBP1020-PA NIBP1020-BPA NIBP1000-KIT BASIC UNIT + PEAK + BATTERY & PEAK BASIC UNIT WITH EKG + PEAK & ALARM + BATTERY, PEAK & ALARM NIBP1020-BPA, 5 ADAPTERS & SOFT CARRYING CASE OPTIONAL ACCESSORIES: • • • • • • • BC20-30111 BC20-40714 BC20-40715 BC20-40716 BC20-40717 BC20-40608 BC20-40602 • BC20-40605 SOFT CARRYING CASE DINAMAP/CRITIKON ADAPTER QUICK DISCONNECT ADAPTER LUER ADAPTER MARQUETTE ADAPTER BULB ADAPTER NIBP FITTING KIT (HP CUFF ADAP
OVERVIEW This section looks at the layout of the NIBP-1010 and gives descriptions of the elements that are present. LCD Graphical Display: Shows Test Data and Configuration Data Male Quick Connect for Pressure Input 6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.
This section looks at the layout of the NIBP-1020 and gives descriptions of the elements that are present. LCD Graphical Display: Shows Test Data and Configuration Data Male Quick Connect for Pressure Input 6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.
KEYS Six tactile-touch keys are provided for system operation: – This key turns the unit off and on. The unit will return to the screen that was active when it was turned off. – In the DISPLAY MODE, these keys toggle the display through the available main screens. In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through the available settings. – On the Main screen, this key sequences through the available NIBP or ECG simulations.
SCREENS MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional). The available screens can be toggled using . PRESSURE ONLY –This screen has a large pressure display, as shown below.
OUTPUT WAVEFORM– This screen shows the output waveform as well as the pressure, peak pressure, and output selection. NOTE This waveform is not intended to be physiologically correct.
PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as, the pressure, peak pressure, and output selection PRESSURE PEAK PRESSURE (Optional) PRESSURE GRAPH 2mmHg per pixel OUTPUT SELECTION Systolic/Diastolic (mean) Adult 120/80 Neonatal (Neo) 70/40 Hypertensive (High) 190/120 Manometer Alarm Test (Optional) (Alternates every 30 sec) Adult 120/80 Mean Heart Rate 93 55 80 94 Respiration Rate (Optional) 35 47 150 80 35 - 47 160 Pacemaker 70 00 80 35 93 10
ECG OUTPUT SCREEN – This screen shows the selected ECG output mode. (NIBP-1020 units only) NOTE: While in this mode, the NIBP simulation does not run. NOTE: NRS ECG output is active during NIBP testing at the rate stipulated for the selected test. The display will resemble the following: LIST OF AVAILABLE OUTPUTS SELECTED OUTPUT NSR NSR NSR NSR Sine Wave Sine Wave Sine Wave Square Wave Square Wave Triangle Wave 11 30 Bpm 60 Bpm 120 Bpm 240 Bpm 10 Hz 60 Hz 100 Hz 0.
BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage level with Built-In Battery Option only. NOTE: It is only an estimate of the battery life remaining. When the level reaches 10%, the BP simulation mode will be disabled; however, the Manometer and ECG (optional) will continue to function. Once the battery level reaches 0%, the unit will automatically turn itself off to avoid damaging the batteries.
SETUP The Setup Mode allows the user to adjust the configuration of the simulator. The Setup screen can be entered using the key. The parameters can be changed by using key to highlight the line and to toggle the available options. The Setup screen can be exited using the key.
SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option to shift test results to compensate for different methods of measuring Oscillometric NIBP by various manufacturers and models of devices under test. CAUTION These adjustments must be used with caution as they will allow the user to adjust the output results to invalid values. These adjustments should only be used to aid in the simplification of testing and with documented controls.
OPERATIONS CONNECTING PRESSURE The NIBP-1010 and NIBP-1020 Blood Pressure Simulators are connected between the blood pressure cuff and the monitor (Device Under Test, DUT). The blood pressure cuff should be disconnected from the DUT and a ‘T’ adapter inserted between the cuff and the DUT. The NIBP simulator is then connected to the open side. Blood Pressure Cuff Monitor (Device Under Test) (DUT) NIBP1010 or NIBP1020 Roll the cuff tightly on itself or around a mandrel.
CONNECTING PATIENT LEADS For the NIBP-1020, with ECG output, test snaps are provided along the sides and are identified by the markings on the overlay.
THEORY OF OPERATIONS A motor and piston assembly is used to generate the output waveforms. A differential pressure sensor is used to measure the pressure of the cuff. The pressure sensor is read by a 16-bit differential Digital to Analog converter. The ECG output is performed by a 12-bit Digital to Analog converter. A second 12-bit Digital to Analog converter is used to generate the respiration waveform.
RUNNING A TEST NIBP: To run an NIBP simulation, the cuff and monitor are connected to the pressure input. Then the measurement is initiated by the monitor and the NIBP-1010 and NIBP-1020 will output the proper waveform based on the cuff pressure. The NIBP output mode can be changed by pressing the Select key. Once the desired operating mode is selected, the output will automatically begin once the correct pressure is detected.
MANUAL REVISIONS Revision # Program # Revisions Made Rev 01 Rev 02 Rev 03 Rev 04 Rev 05 Rev 06 DT7355CA DT7355CA DT7355CA DT7355CF DT7355CG DT7355CI Origination Specification Information Updated Miscellaneous Edits Add Systolic/Diastolic adjustments Output Selection edited Miscellaneous Edits LIMITED WARRANTY WARRANTY: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED.
SPECIFICATIONS NIBP-1010 Blood Pressure RANGE ACCURACY 0-500 mmHg @ 20°C +/- (1% of Reading + .5 mmHg) RATE 80, 94 bpm ACCURACY +/- 1% Physical DISPLAYS ENCLOSURE WEIGHT FACE PLATE OPERATING RANGE STORAGE RANGE LCD Graphical 128 X 64 Pixels 7 x 5 x 4 Inches (177.8 x 127.0 x 101.6 mm) ABS Plastic < 3 Lbs (< 1.
NIBP-1020 Blood Pressure RANGE 0-500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg) RATE 80, 94 bpm (synchronized to ECG) ACCURACY +/- 1% ECG NSR RATE 30,60,120,240 BPM ACCURACY +/- 1% AMPLITUDE 2.75 mV ACCURACY +/- 2% @ Lead II ECG Performance SINE WAVE SQUARE WAVE 10,60,100 Hz 0.125, 2.000 Hz TRIANGLE WAVE 2.000 Hz RATE ACCURACY AMPLITUDE AMPLITUDE ACCURACY +/- 1% 2.
Respiration RATE ACCURACY IMPEDANCE DELTA ACCURACY BASELINE ACCURACY +/- 1% 3.0 ohms +/- 10% 1000 ohms +/- 5% Physical DISPLAYS ENCLOSURE WEIGHT FACE PLATE OPERATING RANGE STORAGE RANGE LCD Graphical 128 X 64 Pixels 7 x 5 x 4 Inches (177.8 x 127.0 x 101.6 mm) ABS Plastic < 3 Lbs (< 1.