User manual

45
The PS-2200 provides the unique feature of a Training Mode to aid the user in practicing
the identification of arrhythmias. The unit will sequence through the arrhythmias, allowing
the user to look at the output on their equipment, identify the arrhythmia and then verify
their conclusion with the correct name shown on the display. The user can select either
manual or timed sequencing, as well as whether the arrhythmias will display in order or
randomly. Subsets of the Arrhythmias can be selected to allow for individualization.
The PS-2200 is placed into TRAINING mode by pressing the category key.
The display will resemble the following:
The Timer, Randomizer and Arrhythmias can be selected by using to
highlight the parameter to change and using to scroll to the desired option.
Then is used to accept the new setting.
Manual*
10 Sec
15 Sec
20 Sec
25 Sec
30 Sec
On
Off*
All*
Subset
* Indicates Default Setting
(See Power Up Settings)
TIMER
RANDOMIZER
TRAINING
GROUPING