Step 5:
Press the attached tip firmly onto the extraction reagent tube containing the
processed sample (threading or twisting is not required). Mix thoroughly by
swirling or flicking the bottom of the tube.
NOTE: Do not use tubes or tips from any other product, including other products
from BD or other manufacturers.
After processing the swab in the extraction reagent, the sample should be analyzed within 30 minutes.
After step 5, choose from the BD Veritor Plus Analyzer workflow option below before continuing to step 6:
BD Veritor Plus Analyzer in
Analyze Now mode
BD Veritor Plus Analyzer in
Walk Away mode
BD Veritor Plus Analyzer with the
BD Veritor InfoScan module
in Analyze Now
mode
in Walk Away
mode
Instructions in
section:
A
B
C
D
A
Using a BD Veritor Plus Analyzer inAnalyze Now mode:
Step 6A: Adding the specimen to the test device
Invert the extraction reagent tube and hold it vertically (approximately one inch
above the sample well).
Gently squeeze the ridged body of the tube, dispensing three (3) drops of the
processed specimen into the sample well.
Excess volume remains for retesting if necessary.
NOTE: Squeezing the tube too close to the tip may cause leakage.
Step 7A: Timing test development
After adding the sample, allow the test to run for
15 minutes
before inserting
the test device into the BD Veritor Plus Analyzer.
During incubation time, turn the
BD Veritor Plus
Analyzer on by pressing the
blue power button once.
NOTE:
If running test under laminar flow hood, cover test device to avoid
inconsistent flow.
Step 8A: Using the BD Veritor Plus Analyzer
The BD Veritor Plus Analyzer will complete a self-test before it is ready for
use. After the self-test the display window shows INSERT TEST DEVICE OR
DOUBLE-CLICK BUTTON FOR WALK AWAY MODE.
INSERT THE TEST DEVICE when the 15-minute assay development time
is complete.
The status of the assay analysis process appears in the display window.
Follow the on-screen prompts to complete the procedure. Do not touch the
Instrument or remove the test device until the result appears.
When analysis is complete, the test result appears in the display window.
Step 9A: Record the Result before removing the test device
ATTENTION: TEST Results are NOT maintained in the display window when the device is removed or if the
BD Veritor Plus Analyzer is left unattended for more than 15 minutes (60 minutes if AC power adapter is connected).