Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
The BD Strive™ system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in such an
environment.
Immunity Test IEC 60601 level
Compliance
level
Electromagnetic
Environment Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV
Contact
± 15 kV air
The BD Strive™ system
should not be aected by
electrostatic discharge that
might occur under normal
conditions of use.
Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
(Continued)
Immunity Test IEC 60601 level
Compliance
level
Electromagnetic
Environment Guidance
Electrical fast
transient/burst
IEC 61000-4-4
+2 kV for power
supply lines
+1 kV for input/
output lines
Not
Applicable
Requirements does not
apply to BD Strive™ system
because it is a battery
powered device.
Power Line Surge
IEC 61000-4-5
±1 kV dierential
mode
±2 kV common
mode
Not
Applicable
Requirements do not
apply to BD Strive™ system
because it is a battery
powered device.
92038.01_IFUCanada.indd 288-289 10/18/18 8:53 AM