User's Guide

Technical Specifications
290
291
Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
Immunity Test IEC 60601 level
Compliance
level
Electromagnetic Environment
Guidance
Voltage
Dips, Short
Interruptions,
and Voltage
Variations
IEC 61000-4-11
<5% U
(>95% dip in U)
for 0.5 cycle
(U is ac mains
voltage prior to
application of the
test level)
Not
Applicable
Requirements do not apply to
BD Strive™ system because it
is a battery powered device.
Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
(Continued)
Immunity Test IEC 60601 level
Compliance
level
Electromagnetic
Environment Guidance
Power Frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz
to 80 MHz
Not
Applicable
Requirements do not
apply to BD Strive™ system
because it is a battery
powered device.