User's Guide

Technical Specifications
292
293
Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
(Continued)
Immunity Test IEC 60601 level Compliance level
Electromagnetic
Environment Guidance
Radiated RF
IEC 61000-4-3
10V/m for
80~2700 MHz,
10V/m 80% AM
at 1 kHZ
10V/m for
80~2700 MHz,
10V/m 80% AM at
1 kHZ
Portable and mobile RF
communications equipment
should be used no closer
to any part of the BD
Strive™ System than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
IEC 60601-1-2
ed 4.0 Clause
8.10
Refer to table
on page 284 for
details
Refer to table
on page 284 for
details
Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
(Continued)
Electromagnetic Environment Guidance (Continued)
Recommended separation distance:
d=1.17 √P 150 kHz to 80 MHz
d=0.35 √P 80 MHz to 800 MHz
d=0.7 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W), according to
the transmitter manufacturer, and d is the recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey
a
, should be less than the compliance level in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
For reference, please see table on page 284 for the compliance level of BD Strive™ system
against common radio transmitters.