User's Guide

Dose accuracy
306
307
1.1 Dose accuracy
The basal dose accuracy was evaluated for three samples. The
following graphs show the typical basal flow rate accuracy of the
system against given time periods. The measurements were taken
at an intermediate basal rate of 40U/day in accordance with
60601-2-24 at controlled room temperature.
Stabilization Period Flow Rate
Hours
Flow Rate (ul/h)
Target Flow rate = 16.67 uL/hr
0
0
10
20
30
40
50
6 12 18 24
IDD flow rate upon initial startup (per IEC 60601-2-24)
Average flow rate during 24 hours period.
8.5 Dose accuracy
92038.01_IFUCanada.indd 306-307 10/18/18 8:54 AM