User's Guide

Prepare BD Strive
91
Step 2. Connect to IDD (continued)
90
h. Tap
Yes
on Remote to connect to IDD. If a communication
error occurs (Section on page 246) move to another room and
try connecting again.
i. Place IDD back into tray to help you hold IDD steady when
filling. While the IDD is starting up, you can begin preparing
the fill syringe. Proceed to step 3.
IMPORTANT
:
For successful pairing, keep IDD and Remote
in close range to each other (Remote above
the IDD). Ensure you are at least 6 feet (2m)
away from other Remotes, IDD’s, or other
equipment that might interfere with the
pairing communication.
Once IDD is connected, you will have 60
minutes to complete setup step.
92038.01_IFUCanada.indd 90-91 10/18/18 8:51 AM