Datasheet

www.Amprobe.com
Data Sheet
Function Range Accuracy
Specifications (at 23 °C ± 5 °C, <75 % R.H. non-condensing), cont.
30XR-A Digital Multimeter
Continuity
Audible Indication 75 Ω ± 25 Ω
Response Time 100 ms
Diode Test
Test Current 1.0 mA (approximate) ± (1.5 % rdg + 3 dgts)
Resolution 0.001 V
Open Circuit Volts 3.0 V typical
Battery Test
Ranges 1.5 V, 9 V ± (3.5 % rdg + 2 dgts)
Resolution 1 mV, 10 mV
Load Test Current 1.5 V range: 150 mA typical
9 V range: 5 mA typical
Non-Contact Voltage Detector (NCV)
AC Volts 70 to 600 V ac
Red LED and Audible Indicator
Overload Protection
AC/DC Voltage 400 mV range: 600 V DC or 600 V AC RMS 15 sec
All other ranges: 600 V DC or 600 V AC RMS
AC/DC Current uA/mA input: F 0.25 A/ 600 V Min. I.R. 30 kA (6.3 x 32 mm)
10 A input F 10A / 600V, Min. I.R. 100 kA, (10 x 38 mm)
(10 A for 4 minutes maximum followed by a 12 minute cooling period)
Resistance 600 V DC or 600 V rms
Continuity 600 V DC or 600 V AC RMS
Diode Test 600 V DC or 600 V AC RMS
Battery Test 600 V DC or 600 V AC RMS