Fuse replacement (mains fuse F1)
In case mains switch pilot lamp (3) does not illuminate after connecting the
Machinery-Tester MT204-S to mains outlet and switching on the mains
switch and neither the LC-Display (16) shows any indication, it is very likely
mains fuse (2) to be blown.
To replace the fuse proceed as follows:
1) Open the fuse holder 1 (2) above the ON/OFF switch by using an
appropriate screwdriver
2) Remove the defective fuse and replace it with a new one (T 1 A /
250 V, 5 × 20 mm)
3) Close the fuse-holder again
Fuse replacement (fuse F2 for RPE10A, ZI, RISO and HV
functions)
Internal fuse F2 (FF 12.5 A / 500 V, 6.3 × 32 mm) has blown if:
Text "FUSE (F2) 12.5A" appears on the LC-Display in RPE10A, RISO or HV
function
Measurement values in ZI function are very low (ZI lower than 0.05 or
IPSC higher than 4.6 kA)
Check the fuse also if "FUSE (F2/F3)" appears on the LC-Display in RPE0.2A
or RPE10A function
To replace the fuse proceed as follows:
1) Unlock corresponding fuse holder (4) by using an appropriate
screwdriver
2) Remove the defective fuse and replace it with a new one (FF 12.5 A
/ 500 V, 6.3 × 32 mm).
3) Lock the fuse holder again.