Technical data
General data
Display Graphic LC-Display, 128 x 64 dots
Limit value display Optic and acoustic
Limit value setting Within measurement range (in functions
RPE, Zl and RISO)
Memory Approx. 2000 memory locations, 3 levels
(customer, machine, measurement location),
additional measurement No. is created.
Interface (USB1) USB 2.0 device, USB interface to PC
Interface (USB2) USB 2.0 host, for barcode reader, keyboard
or USB memory stick
Requirements on USB
memory stick FAT12, FAT16 or FAT32 with a sector size of
512 Byte
Working temperature range 0 to 40°C (accuracies given in technical data
refer to this range)
Storage temperature range -10°C ... +50°C
Allowed humidity range 10% ... 85% relative humidity (without
condensation)
Height above sea level up to 2000 m
Mains power supply 230 V ± 10%, 50 Hz
Max. power consumption 230 VA
Installed fuses:
Mains fuse (F1) T1 A / 250 V, 5x20 mm
For functions
RPE 10A, RISO, HV (F2) FF12.5 A / 500 V, 6.3x32 mm
RPE 0.2A (F3) FF1 A / 500 V, 6.3x32 mm
Constructed according to EN 61010-1 / EN 61557 (parts 2, 3, 4)
Over-voltage category:
Mains inlet CAT II / 300 V
Measurement input CAT IV / 300 V, CAT III / 440 V
Measurement output (HV) CAT I / 1300 V
Measurement input / output is isolated with basic insulation against
ground. Maximal output voltage (open circuit, mains over-voltage) is
1300 V.
Degree of pollution 2
Protection class I
IP protection IP 40
Dimensions (W × L × H) 345 x 320 x 170 mm with handle
Weight Approx. 7 kg