User Manual

15
Normal mode
for the MAINS circuit:
Connect the test lead L1/L2/L3 to the 3-phase mains circuit by using probes
and/or alligator clips. If the meter indicates a clockwise movement, the
phases connected to L1, L2 and L3 of the meter are L1, L2 and L3 (also
known as R, S and T), respectively. If the meter indicates a counter-clockwise
movement, swap any two connects between the meter and phases. Then
retest. Connecting the above mentioned L1, L2 and L3 of a motor and that
of the MAINS respectively should get a clockwise motor movement.
Using the Beeper feature on
& mode:
Press and hold the REC button while turning the rotary switch to
to enable the beeper, the display shows “Enbp”. When making rotary
measurement, if the rotary indication segments indicate a clockwise
movement, the beeper sounds a single long beep per segment cycle. If the
rotary indication segments indicate a counter clockwise movement, the
beeper sounds 3 short beeps per segment cycle.
Auto Power Off
The meter turns off if there is no button pushed, rotary function switch
operation for 32 minutes and/or no specified activities below, where
applicable:
1.) Significant measuring readings of above 8.5% of ranges
2.) Non-OL readings for Resistance, Continuity or Diode function
3.) Non-zero readings for Hz function
4.) Significant movement indication as in Phase Rotation functions
The meter will not enter auto power off mode when under normal
measurements. To turn the meter back on, press the SELECT button and
release, or turn the rotary switch off and on to restart the meter.