User's Manual

V1.0M122 PULSE OXIMETER OPERATORS MA NUAL
17
are divided by built-in module and can not be changed by user.
Assignment of priority:
High Medium Low
Paramter
SpO
2
PR
Value
Red Yellow
Alarm lamp
Flashing with yellow Flashing with yellow
Lamp
Frequency
1.5Hz 0.5Hz
Audiblesound
Di- Di – Di ----- Di - Di Di - Di - Di Di
Alarm cycle
3 s 5 s 20 s
Alarm info
SpO
2
too high/low PR too high/low Probe off/Finger out
Notes:
1. The alarm sound will go on until alarm disappears or is turned off.
2. After silencing the alarm, the corresponding indicator will indicate this.
3. The power low alarm: the corresponding indication lamp will be flashing with a red
frame.
AUDIBLE ALARM INHIBITION/OFF:
Short press the
button to silence the audible alarm for 2 minutes, the audible alarm
indicator will be displayed as
together with the countdown from 120s to 0sshort
press it againyou can cancel alarm inhibition; Long press the
button, the audible
alarm indicator will be displayed as
which indicates the audible alarm off. And then
long press the
button, you can activate the audible alarm again.
Warnings!
When an alarm occurs, check patients’ conditions immediately.
z Check which parameter is alarming or which alarm is going on.
z Check patient’s condition.
z Search for the source of alarm.
z Make the alarm mute if necessary.
z Check the alarm when no warning.
6.2 Data Management
From the main menu screen, select and enter the “Data Manage” screen, refer to fig.6-3.