APPENDIX 4 Thank you for choosing the Biomega BioClave steam sterilizer. HYDRAULIC DRAWING Your steam sterilizer has been CE certified and designed with durability, reliability, and safety in mind. It is your responsibility to install this instrument in conformance with local electrical codes. This manual contains important operating and safety information. Please read and understand the contents of this manual prior to operating this instrument.
APPENDIX 3 TABLE OF CONTENTS 1. GENERAL ------------------------------------------------------------ 1 ELECTRICAL DRAWING 2. TECHNICAL SPECIFICATIONS----------------------------------- 2 3. PACKING CONTENT ----------------------------------------------- 3 V1 4. INSTALLATION ---------------------------------------------------- 4 ¡ 21V V4 ¡ 9V 5. CONTROL PANEL --------------------------------------------------- 5 6 .
1. General The sterilizer described in this manual is intended for the sterilization of research tools. It operates automatically with 134°C and 121°C sterilization temperatures. This sterilizer is in compliance with the European Directive 93/42/CEE and it has been produced in accordance with the EN 13060. In addition the chamber has been ASME certified.
2. Technical Specifications (1)Chamber Dimensions: 9in. x 14.2in. / 230mm x 360mm Pressure psi WRAPPED 30 134°C (2)Rated Voltage: AC110V-120V or 220-240V, 50-60Hz (3)Nominal power: 1600W (120V) 1250W(220V) (4)Sterilization Temperature: 121°C / 134°C (5)Main Fuses: T20A/250V 15 (6) Capacity of the distilled water tank: Approx 2.5L (water at level MAX) Time (min) 0 (7)OperatingEnvironment: 5 - 40°C (8)External Dimensions: 17.5in. (width) x 16.7in. (height) x 23in.
3.
APPENDIX 2 4. Installation 4 MAXLOAD (kg) 30 Total time (min.) 134 Holding time (min.) Pressure (PSI) PROGRAM Temperature (°C) DIAGRAMS OF THE STERILIZATION PROGRAMMES TYPE 12~30 5.00 Unwrapped solid material UNWRAPPED * Ensure that the sterilizer is installed with 2.5in. (10cm) ventilation space on all sides of the sterilizer, and 5 in. (20cm) on top side. The clearance required to open the door is 15.5in. (40cm). * The sterilizer should be placed on a level worktable.
12. Error Codes * Verify all samples are not sealed, capped or closed. * Don't overload the trays above the stated limit (see Appendix 1). * Don't stack the trays one above the other or put them in direct contact with Code the walls of the sterilization chamber. * When handling the trays, always use the tray removal tool.
10.6 Replace fuse 6.6 After the samples are loaded, close and lock the door by turning the door handle clockwise until it stops. 1). Switch off the power. 2). Push inward with a flat head 3). Pull out the fuse holder by screw driver, then unscrew the fuse hand. holder ( counter clockwise). Caution: You must turn the door handle to the maximum position, otherwise the machine will alarm and an error message will be displayed during the cycle. 6.7 Starting the sterilization program. 4).
10.5 Replacement of the door seal 7. Draining the water tanks 10.5.1 Fully open the door. 10.5.2 Remove the door seal carefully by pulling it by hand away from the door 10.5.3 Clean the new door seal carefully with a soft cloth saturated with distilled water. 10.5.4 Moisten the new seal with medical disinfectant . 1. Connect the supplied drain hose to 2. Set the drain valve to the open the drain valve by pressing it on firmly. position by turning it counter clockwise 10.5.
10.3 Cleaning the inner-chamber, weekly - Remove the trays and tray rack from the inner-chamber - Wipe the inner chamber with a soft cloth saturated in distilled water - Apply the same procedure to the trays and tray rack. 10.4 Door adjustment Below are the instructions for the door adjustment. This should only be performed if the door is not providing the proper seal or if the door seal has just been replaced. 10.4.
9. Printer (Optional) 9.1 If you have purchase the BioClave printer device, you can connect the printer cable to the back of the sterilizer. 9.2 Connect the printer power cord to the socket with the proper electrical specifications. 9.3 Power on the sterilizer and press the “sel” button. 9.4 After the sterilization cycle has completed, the printer will begin printing the details of the sterilization cycle (see below). 10 Maintenance 7.