Datasheet

Serial no.:
Date:
Performed by:
14400033
24.4.2015
Koprivec, Petkovšek, Šuštarši
BENNING IT 115
KALIBRIERZERTIFIKAT/CALIBRATION CERTIFICATE
Page 1 of 3
PASS
Outlook, BAT indication, keys, communication1.
Nr.:
No.:
Eingang
Input
Unterer
Grenzwert
Low limit
Anze ige
Reading
Oberer
Grenzwert
High limit
Insulation
50 V 60 V
500 V 600 V
52
527
3.
Insulation 0 M 0,00 M 0,03 M0,004. 50 V
10 M
9,47 M 10,53 M
100 M
90,0 M 110,0 M
10,18
104,8
0 0,00 0,03
1 0,94 1,06
0,00
0,99
Continuity ± 200 mA
R low
7.
0 V
0 V 2 V
230 V
223 V 237 V
AC voltage TRMS
1
230
2.
50 V
500 V
Unsicherheit
Uncertainty
1 V
3 V
1 V
1 V
Insulation
1 M 0,92 M 1,08 M
1,00
5.
10 M
9,47 M 10,53 M
100 M
95 M 105 M
10,14
101
900 M
810 M 990 M887
19 18,40 19,60
1900 1805 1995
19,05
1950
50 Hz
49,8 Hz 50,2 Hz50,0 0,1 Hz
frequ ency
R ISO
UisoOut - 5 V
UisoOut + 5 V
52
1 V
10 MUiso output
Uiso displayed
Uiso output
Uiso displayed
UisoOut - 19 V UisoOut + 19 V
525
3 V
1000 V 1200 V
1054
1000 V
6 V
Uiso output
Uiso displayed UisoOut - 35 V UisoOut + 35 V
1050
6 V
INS
500 VR ISO
0,1 M
0,07 M 0,13 M0,10
19 17,6 20,4
19,0
19 17,6 20,4 19,1
2 200 mA 250 mAPASSUbat=7 V
R+
R-
0,01 M
0,06 M
1,0 M
0,01 M
0,06 M
1,0 M
10 M
0,01 M
0,01
0,01
0,11
11
0,1
0,1
Funktion / UBat = 7,5 V
Function / Ubat = 7,5 V
10 M
1 M
900 M
810 M 990 M
865 10 M
190 M
180,5 M 199,5 M
189
1,1 M
1 M
0,92 M 1,08 M1,00 0,01 MInsulation6. 1000 VR ISO
Datum der Inbetriebnahme:________________________gültig bis:______________________*
BENNING Kalibrierintervall-Empfehlung: 12 Monate
*Das Gültigkeitsdatum berechnet sich aus dem Datum der Inbetriebnahme zuzüglich der
Kalibrierintervall-Empfehlung.
230 V
223 V 237 V230 1 V
U
L-N
U
L-PE
U
N-PE
0 V 0 V 2 V
230 V 223 V 237 V
0
230
1 V
1 V
230 V
223 V 237 V230 1 V
U
L-N
U
L-PE
U
N-PE
change L-N
Date Placed In Service: ____________________ Due Date: ________________________*
BENNING Recom mended Cal Interval: 12 months
* The due date may be established (by the customer) by adding the "Recommended Cal
Interval" to the "Date Placed In Service."
595
installation tester
BENNING IT 105 – testing
of electrical installations
according to IEC 60364
• unbeatable price/performance ratio
• easy operation, all measuring functions can be
selected directly by means of a rotary switch
• large LC display with illumination
• supplied in a hard-top case, test probe with „TEST“
key, set of measuring leads/ alligator clips,
test cable with shock-proof plug, carrying strap,
6 x 1.5 V batteries (IEC LR6/ type AA) and
calibration certificate
duspol@benning.de · Tel.: +49 / (0) 2871 / 93-111
www.benning.de
VDE 0100
NEW!
IT 105
VDE 0100 Installation tester
multifunction
incl.
calibration
certificate

Summary of content (2 pages)