- Then, insert the batteries into the battery compartment at the provided places (please ob-
serve correct polarity of the batteries).
- Lock the battery compartment cover into place on the bottom part and tighten the screw.
See gure 9: Battery/ fuse replacement
m
Make your contribution to environmental protection! Do not dispose of
discharged batteries in the household garbage. Instead, take them to a
collecting point for discharged batteries and special waste material. Please
inform yourself in your community.
9.5 Fuse replacement
c
Before opening the BENNING PV 1-1 ensure that all the test leads have been
disconnected from the BENNING PV 1-1. Danger of electric shock!
The 4 mm test sockets K + L of BENNING PV 1-1 are protected against overload by means of
one built-in fuse (500 mA, 1000 V, F, D = 6.3 mm, L = 32 mm) (P.no.749771).
Proceed as follows to replace the fuse (see Figure 10):
- Switch the BENNING PV 1-1 off.
- Put the BENNING PV 1-1 face down and unscrew the screw of the battery compartment
cover.
- Lift off the battery compartment cover (in the area of the housing slots) from the bottom part
of the battery compartment.
- Laterally lift one side of the defective fuse off the fuse holder by means of a slotted screw-
driver.
- Completely remove the defective fuse from the fuse holder.
- Insert the new fuse. Only use fuses of the same nominal current, nominal voltage, nominal
breaking capacity, tripping characteristic and dimensions.
- Lock the battery compartment cover into place on the bottom part and tighten the screw.
See gure 10: Battery/ fuse replacement
Note:
The PV test sockets
9
+ J of BENNING PV1-1 panel shorting circuitry are protected by a
15 A, 1000 V Solar fuse. This fuse is not operator replicable. If this fuse blows then the
BENNING PV 1-1 will indicate an error before the panel is shorted, the BENNING PV1-1 must
be returned for service.
9.6 Calibration
To maintain the specied precision of the measurement results, the instrument must be recali-
brated at regular intervals by our factory service. We recommend a recalibration interval of one
year. Send the appliance to the following address:
BENNING Elektrotechnik & Elektronik GmbH & Co. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt
9.7 Spare parts
Fuse 500 mA, F, 1000 V, breaking capacity ≥ 500 A, D = 6.3 mm, L = 32 mm, P.no. 749771
10. Environmental notice
At the end of the product’s useful life, please dispose of it at appropriate collection
points provided in your country.