Operation Manual

09/ 2013
BENNING CM 6
11
IP 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 - first index). No protec-
tion against water, (0 - second index).
- Operating temperature and relative humidity:
For operating temperature from 0 °C to 30 °C: relative humidity less than 80 %
For operating temperatures from 31 °C to 40 °C: relative humidity less than 75 %
For operating temperature from 41 °C to 50 °C: relative humidity less than 45 %
- Storage temperature: The BENNING CM 6 can be stored at any tempera-
ture in the range from - 20 °C to + 60 °C (relative humidity from 0 to 80 %).
The battery should be taken out of the instrument for storage.
7. Electrical specications
Note: The measuring precision is specified as the sum of
- a relative fraction of the measured value and
- a number of digits (counting steps of the least significant digit).
This specified measuring precision is valid for temperatures in the range from
18 °C to 28 °C and relative humidity less than 80 %.
7.1 Direct voltage ranges
The input resistance amounts to 1 MΩ
Measuring range Resolution
Measuring precision Overload protection
400 V 0,1 V ± (0,7 % of the measuring value + 2 digit) 750 V
eff
1000 V 1 V ± (0,7 % of the measuring value + 2 digit) 750 V
eff
7.2 Alternating voltage ranges
The input resistance amounts to 1 MΩ in parallel 100 pF.
Measuring range Resolution
Measuring precision
*1
within the
frequency range 50 Hz - 500 Hz
Overload protection
400 V 0,1 V ± (1 % of the measuring value + 5 digit) 750 V
eff
750 V 1 V ± (1 % of the measuring value + 5 digit) 750 V
eff
*
1
The measured value is obtained by mean value rectification and displayed
as r.m.s. value. Its calibration is matched to a sinus-shaped curve type.
7.3 Alternating current ranges
Measuring range Resolution
Measuring precision
*1
within the
frequency range 50 Hz - 60 Hz
Overload protection
60 A 0,1 A ± (1,9 % of the measuring value + 7 digit) 1000 A
eff
400 A 0,1 A ± (1,9 % of the measuring value + 5 digit) 1000 A
eff
1000 A 1 A ± (1,9 % of the measuring value + 5 digit) 1000 A
eff
within the frequency range 61 Hz - 400 Hz
400 A 0,1 A ± (2,5 % of the measuring value + 7 digit) 1000 A
eff
1000 A 1 A ± (2,5 % of the measuring value + 7 digit) 1000 A
eff
*
1
The measured value is obtained by mean value rectification and displayed
as r.m.s. value. Its calibration is matched to a sinus-shaped curve type.
The stated precision is specified for conductors that are centrally clamped
by the current probe (see Fig. 4 alternating current measurement). For
conductors that are not centrally clamped, an additional error of 1 % of the
display value needs to be taken into account.
7.4 Resistance measuring range and acoustic continuity testing
Overload protection: 600 V
eff
Measuring range Resolution Measuring precision
Max. idling voltage
400 Ω 0,1 Ω ± (0,7 % of the measuring value + 3 digit) 3 V
The built-in buzzer sounds in the case of a resistance R less than 30 Ω.
7.5 Frequency ranges
Overload protection: 1000 A
eff
Measuring range Resolution Measuring precision
400 Hz 1 Hz ± (0,1 % of the measuring value + 2 digit)
Minimum input frequency: 20 Hz
Minimum sensitivity: 3 A
eff