Operation Manual
BENNING IT 130 short instructions
- 13 -
2 Measurements
2.1 Null balance (compensation) of the test cables
1. Set function
R LOW
Select R LOW or CONTINUITY.
Short-circuit the test cables.
Shorted test leads
3. Press the key .
4. Press the key CAL (HELP).
After performing test leads compensation first measured value and then 0.00 is displayed.
Successful compensation is displayed by means of the symbol in the "R LOW" and
"CONTINUITY" function.
Resistance of the test cable before compensation
Resistance of the test cables after compensation