Operation Manual

BENNING IT 130 short instructions
- 8 -
The test cable resistance for the low-impedance measurement / continuity test has
not been compensated.
The test cable resistance for the low-impedance measurement / continuity test has
been compensated.
High resistance to earth of test probes. Results may be impaired.
The current is too low for the accuracy specified. This might result in incorrect
measuring results. Please check in the current clamp settings whether the accuracy
of the current clamp can be increased.
Measured signal is out of range (clipped). Results are impaired.
Simple error in the IT network.
Fuse F1 is broken.
Sound warnings
Continuous
sound
Warning! Dangerous voltage on the PE terminal is detected.
Evaluation of the measuring result
Measurement result is inside pre-set limits (green LEDs).
Measurement result is out of pre-set limits (red LEDs).
Measurement is aborted. Consider displayed warnings and messages.
1.5 Selecting measuring functions
Rotary switch for selecting the
measuring function
Selects sub-function
Selects value of parameter / limit
Selects parameter / limits
Starts measurement