Operation Manual
17
200 nF 100 pF
± (1.9 % of the measured value + 8 digits)
2 µF 1 nF
± (1.9 % of the measured value + 8 digits)
20 µF 10 nF
± (1.9 % of the measured value + 8 digits)
200 µF 100 nF
± (1.9 % of the measured value + 8 digits)
2 mF 1 µF
± (1.9 % of the measured value + 8 digits)
< 10 digits in case of uctuating indication
7.8 Frequencyranges(BENNING MM 1-2/ 1-3)
Overload protection for frequency measurements: 600 V
eff
Minimum pulse width > 25 ns; operating cycle limitation > 30 % and < 70 %
Measuring
range
Resolution Measuring accuracy for 5 Veff max. Sensitivity
2 kHz 1 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
20 kHz 10 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
200 kHz 100 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
2 MHz 1 kHz
± (0.01 % of the measured value + 1 digits)
> 2 < 5 V
eff
20 MHz 10 kHz
± (0.01 % of the measured value + 1 digits)
> 2 < 5 V
eff
7.9 Temperature ranges °C (BENNING MM 1-3)
Temperature measurements (BENNING MM 1-3) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
eff
Measuring range Resolution Measuring accuracy
-20 °C ~ 0 °C 1 °C ± (2 % + 4 °C)
1 °C ~ 100 °C 1 °C ± (1 % + 3 °C)
101 °C ~ 500 °C 1 °C ± (2 % + 3 °C)
501 °C ~ 800 °C 1 °C ± (3 % + 2 °C)
7.10 Temperature ranges °F (BENNING MM 1-3)
Temperature measurements (BENNING MM 1-3) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
eff
Measuring range Resolution Measuring accuracy
-4 °F ~ 32 °F 1 °F ± (2 % + 8 °F)
33 °F ~ 212 °F 1 °F ± (1 % + 6 °F)
213 °F ~ 932 °F 1 °F ± (2 % + 6 °F)
933 °F ~ 1472 °F 1 °F ± (3 % + 4 °F)
8. Measuring with the BENNING MM 1‑1/ 1‑2/ 1‑3
8.1 Preparing the measurement
Operate and store the BENNING MM 1-1/ 1-2/ 1-3 at the specied storage and
operating temperatures only!
Do not permanently expose the device to sunlight.
- Check stated nominal voltage and nominal current on the safety measuring
leads. Nominal voltage and current of the enclosed safety measuring leads
comply with the respective values of the BENNING MM 1-1/ 1-2/ 1-3.
- Check insulation of the safety measuring leads. If the insulation is damaged,
the safety measuring leads must be replaced immediately.
- Check the safety measuring leads for continuity. If the conductor in the
safety measuring lead is interrupted, replace the safety measuring leads
immediately.
- Before selecting another function by means of the rotary switch 8, discon-
nect the safety measuring leads from the measuring point.
-
Strong sources of interference in the vicinity of the BENNING MM 1-1/ 1-2/ 1-3
might involve unstable readings and measuring errors.
8.2 Voltage and current measurement
Do not exceed the maximum permitted voltage with respect to
earth potential! Electrical danger!
The highest voltage which may be applied to the
09/ 2013
BENNING MM 1-1/ 1-2/ 1-3