Operation Manual

18
- COM jack J
- jack for V, Ω, , Hz 9
- jack for the 10 A range K (BENNING MM 1-2/ 1-3)
of the BENNING MM 1-1/ 1-2/ 1-3 against ground is 1000 V.
Electrical danger!
Maximumcircuitvoltage fora measured currentof 500V!In
caseoffusetriggeringatvaluesabove500V,damagesofthe
device might be involved. A damaged device might represent
an electrical hazard!
8.2.1 Voltage measurement
- Select the desired function (V AC) or (V DC) by means of the rotary switch
8 of the BENNING MM 1-1/ 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack J of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for V, Ω, , Hz 9 of the
BENNING MM 1-1/ 1-2/ 1-3.
- Bring the safety measuring leads into contact with the measuring
points and read the measured value on the digital display of the
BENNING MM 1-1/ 1-2/ 1-3.
See gure 2: DC voltage measurement
See gure 3: AC voltage measurement
8.2.2 Current measurement (BENNING MM 1-2/ 1-3)
- Select the desired range and function (A AC) or (A DC) by means of the
rotary switch 8 of the BENNING MM 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack J of the
BENNING MM 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for the 10 A range K (DC
or AC currents up to 10 A) of the BENNING MM 1-2/ 1-3.
- Bring the safety measuring leads into contact with the measuring
points and read the measured value on the digital display of the
BENNING MM 1-2/ 1-3.
See gure 4: DC current measurement (BENNING MM 1-2/ 1-3)
See gure 5: AC current measurement (BENNING MM 1-2/ 1-3)
8.3 Resistance measurement
- Select the desired function (Ω) by means of the rotary switch 8 of the
BENNING MM 1-1/ 1-2/ 1-3
- Connect the black safety measuring lead to the COM jack J of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for V, , , Hz 9 of the
BENNING MM 1-1/ 1-2/ 1-3.
- Bring the safety measuring leads into contact with the measuring
points and read the measured value on the digital display of the
BENNING MM 1-1/ 1-2/ 1-3.
See gure 6: Resistance measurement
8.4 Diode test
- Select the desired function ( ) by means of the rotary switch 8 of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack J of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for V, , , Hz 9 of the
BENNING MM 1-1/ 1-2/ 1-3.
- Bring the safety measuring leads into contact with the diode connec-
tions and read the measured value on the digital display of the
BENNING MM 1-1/ 1-2/ 1-3.
- For a standard Si diode applied in conduction direction, a conduction
voltage between 0.400 V and 0.900 V is displayed. „000“ indicates a short-
circuit inside the diode, „OL“ indicates an interruption inside the diode.
- For a diode applied in reverse direction, „OL“ is indicated. If the diode is
defective, „000“ or other values are indicated.
See gure 7: Diode test
8.5 Continuitytestwithbuzzer
- Select the desired function ( ) by means of the rotary switch 8 of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack J of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for V, , , Hz 9 of the
09/ 2013
BENNING MM 1-1/ 1-2/ 1-3