Operation Manual
08/ 2012 11 / 58
BENNING ST 750 A
<Printer>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
7.3.1 System Data
The device version and serial number are listed under <Setup> / <System Settings> /
<System Data>. The duration of the individual test procedures (standard = 5 sec.) is determined by
the test time. The safety shut-off terminates a test procedure (standard = 30 sec.).
Thecontactsurfacesbetweenconnectorshaveonlylowtransitionresistances.Todenethetransi-
tionresistance,herethe assumable resistance value is entered (standard = 1Ω). Moreover,it is
denedwhetheraprinterorabarcodescannerisconnectedtotheserialinterface.
Mains pole reversal ON: On the test socket, L and N are interchanged.
Buzzer ON: Acoustic signal in case of faults
Short-circuit test ON: Whenstartingthetest,rstcheckforshort-circuit.
Standard test procedures:
Automatic test procedures complying with DIN VDE
0701/0702/0752 are activated.
Custom test procedures:
Customized test procedures are activated. Customized test pro-
cedures are stored on the SD memory card.
Test repetition ON:
This mode allows measurements of several measuring points by
means of the test cable (probe). It can be used for measuring
RPW, RISO (protection classes II/III), Icont, Ileak and IPLeak.
The corresponding maximum and minimum value of all measure-
ments will be stored.
Database backup ON:
The device automatically creates up to 5 backups for each data-
base, e.g. Database name_backup.001.
7.3.2 Factory Settings
Select <Setup> / <System Settings> / <Factory Settings> to reset all limiting values to the factory
settings.Conrmyourentrywith<Yes> / <No>!
08/ 2010 10 / 58
BENNING ST 750
RPE data (resistance measurement of the PE conductor):
RPE VDE 701-702 = resistance for a cable length of 5 m and 1.5mm²
RPE VDE 751 = resistance for a cable length of 5 m and 1.5mm²
RPE per 7.5 m line length = additional resistance per 7.5 m line length
Line length = total length of the device supply line
Cross-section = cross-sectional area of the device supply line
RInsu data (insulation measurement):
RInsu-1 VDE 701-702 = insulating resistance of primary to PE
RInsu-1 VDE 701-702 heating elements = insulating resistance of primary to PE
RInsu-2 / RInsu-3 VDE 701-702 = insulating resistance of secondary to PE / primary to secondary
RInsu-1 VDE 701-702 Ua (SELV/PELV) = insulating resistance of primary to PE (SELV devices)
RInsu-1 VDE 751 = insulating resistance primary to PE
RInsu-2 VDE 751 = insulating resistance of secondary to PE
RInsu-3 VDE 751 = insulating resistance of primary to applied part
Testing voltage for primary to PE = testing voltage for primary to PE
Testing voltage for secondary to PE = testing voltage for secondary to PE
Testing voltage for primary to secondary = testing voltage for primary to secondary
IAL data (alternative leakage current measurement):
IPE VDE 701-702 protection class I = PE alternative leakage current
ILeak VDE 701-702 protection class II = alternative leakage current of accessible parts
ILeak VDE 751 protection class I = PE alternative leakage current
ILeak VDE 751 protection class I = alternative leakage current of accessible parts
Patient leakage current BF = alternative patient leakage current of applied part, type BF
Patient leakage current CF = alternative patient leakage current of applied part, type CF
ILeak data (leakage current measurement):
IPE VDE 701-702 protection class I = PE leakage current
ILeak VDE 701-702 protection class II = leakage current of accessible parts
ILeak VDE 751 protection class I = PE leakage current
ILeak VDE 751 protection class I = leakage current of accessible parts
Patient leakage current BF = patient leakage current of applied part, type BF
Patient leakage current CF = patient leakage current of applied part, type CF
ICont data (contact current measurement):
ICont VDE 701-702 = contact current to PE
Functional data (functional measurement):
IPE VDE 701-702 = PE leakage current (differential current measurement)
ICont VDE 701-702 = contact current to PE (differential current measurement)
IPE VDE 751 = PE leakage current (differential current measurement)
ICont VDE 751 = contact current to PE (differential current measurement)
Ua data (measurement on PELV / SELV devices):
Max. output voltage = max. output voltage for PELV / SELV devices
Cable data (cable measurement):
Line length = simple line length
Cross-section = cross-sectional area of the conductor
Conductor qty. = conductor quantity (for total length)
7.2 System Settings
Select <Setup> / <System Settings> to modify and permanently save the baud rate, parity, data
bits and stop bits e.g. for barcode scanners, RFID readers, PCs and printers.
see chapter 16 to 18
<Barcode Scanner>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
<RFID Reader>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
<PC>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
<Printer>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
7.2.1 System Data
The device version and serial number are listed under <Setup> / <System Settings> /
<System Data>. The duration of the individual test procedures (standard = 5 sec.) is determined by
the test time. The safety shut-off terminates a test procedure (standard = 30 sec.).
Thecontactsurfacesbetweenconnectorshaveonlylowtransitionresistances.Todenethetransi-
tionresistance,herethe assumable resistance value is entered (standard = 1Ω). Moreover,it is
denedwhetheraprinterorabarcodescannerisconnectedtotheserialinterface.
Mains pole reversal ON: On the test socket, L and N are interchanged.
7.3.3 Language Setting
In the <System Settings> menu, press the <Language Settings> button to select the desired
language. Then, press <Save>.
7.3.4 Entering a Password
AlostpasswordcanonlyberesetbytheBenningServiceteam(seechapter24,"Pro-
ductSupport")!
The expert setting function can only be accessed by means of a password. Upon delivery, the de-
viceisprovidedwiththepassword"Benning".ThepersonresponsiblefortheBENNINGST750A
ApplianceTestershalldeneanewcondentialpassword.
To dene a new password, subsequently press <Setup> / <System Settings> / <Password
Entry> / <Input>. First, the current password is requested. Then, the new password will be as-
signed, compared and saved with <OK>.