Operation Manual

08/ 2012 14 / 58
BENNING ST 750 A
Pressthe<▼>buttontoopenthe"calendarview"toselectthemonthandtheyeardirectly.Select
themonthfromthelistandchoosethecorrespondingyearbymeansofthearrowbuttons(<◄>,
<►>).Tosetthetime,select"HH":"MM":"SS"respectivelyandsetthevaluebymeansofthe
arrowbuttons(<◄>,<►>).
7.7 Calibration / System Calibration
Probe calibration (compensation of the test cable): Depending on the test cable used, the probe
should be calibrated again. This calibration is necessary, because the resistance of the test ca-
ble will be taken into consideration for the measuring result of the protective conductor resistance
measurement! The calibration data will be saved permanently in the device.
Gotothe"Setup Selection"menuandpress<Calibration>.
Connecttheconnectingcable(testcable)totheyellow"Sonde/Probe"jackandtothegreen"Kal./
Cal."jack.Press<Probe Calibration>andconrmthemessagewith<Yes>. Now, probe calibra-
tion has been done successfully.
A ZERO balance / calibration is necessary, if in standstill or with a missing signal the displayed
measured values are not approximately at 0 (e.g. 0.000 mA). For this, go to the start menu and
press <Setup>, <Calibration>. Then, press <ZERO Balance> and start it by pressing <Yes>.
During calibration, no connecting cables of the test sample or other cables must be connected to
the device!
The device shall be calibrated at the manufacturer's in certain time intervals (see chap-
ter23"Maintenance/Calibration").