Operation Manual

08/ 2012 35 / 58
BENNING ST 750 A
11.3 Test Procedure according to VDE 0751-1
The DIN VDE 0751-1 standard applies for tests of medical electrical devices, medical electrical
systems or of parts of those devices or systems complying with the DIN EN 60601-1 (VDE0 750 Part
1) standard which are carried out prior to commissioning, inspection, maintenance or during periodic
testing in order to evaluate the safety of those devices or systems or their parts. For devices which
are not built according to DIN EN 60601-1 (VDE 0751 Part 1), this standard can be applied anal-
ogously under consideration of the safety standards applying to the manufacturing of the device.
(taken from DIN VDE 0751-1:2001-10, section 1.1)
Thetestsmustbecarriedoutbyqualiedpersonnel.Thequalicationmustcompriseprofessional
education, knowledge and experience as well as knowledge of the applicable standards and local
provisions. Personnel evaluating safety must be able to detect possible consequences and dangers
that might be caused by devices not complying with the requirements.
(taken from DIN VDE 0751-1: issue 2008-08, section 4.1)
Prior to testing, accompanying documents have to be checked for the manufacturer's recommenda-
tions regarding maintenance and repair, including conditions and precautionary measures.
In the current issue of the DIN VDE 0751-1 standard, the measurement of the insulating resistance
is required, if it is considered to be convenient. Insulation measurement must not be carried out,
if it is excluded in the accompanying documents according to the manufacturer's specications.
Limiting valuesfor the insulating resistance are not speciedinthe DIN VDE 0751-1 standard
(issue 2008). Here, it is possible to refer to recommendations of the manufacturer or to previously
measured values. The DIN VDE 0751-1 standard recommends the following order of the tests to
be carried out:
According to DIN VDE 0751-1 following sequence of tests to be recommended: