3. Install System 1 Ranger Pro plugin build 30XXX or later.
4. Add a compatible wireless gateway to the System 1 configuration.
5. Enable data collection.
6. (If needed)Wait for the Configuration Out-of-Date message to display.
7.
(If needed)In the General properties tab, click Synchronize . Click Save.
Existing System 1 Installations
If needed, first update System 1 and apply any hot fixes or service packs.
To upgrade the Ranger Pro plugin on System 1:
1. Back up your System 1 configuration.
2. Open the System 1 database on which the Ranger Pro plugin is running.
3. Disable data collection.
4. Close the database.
5. Use Windows Add or Remove Programs to uninstall the existing System 1 Ranger Pro
plugin.
6. Download the compatible Ranger Pro plugin from the Bently Nevada Flexera licensing
server.
7. Install the plugin.
8. Open the System 1 database.
9. Enable data collection.
10. (If needed)Wait for the Configuration Out-of-Date message to display.
11.
(If needed)In the General properties tab, click Synchronize . Click Save.
4.8 Mount Devices
Tools Required
To mount Ranger Pro devices, you need:
l Ranger Pro C-spanner and flat wrench. Available as part of the installation kit, described in
the Ranger Pro Datasheet (document 125M5237), which is available from Bently.com.
You also need:
l Spot facing tool 40 mm (1 ½ inches) diameter
l Steel wire brush
l Drills and thread-taps
l Marker pen
l Medium strength thread locking compound, for example, Loctite Blue 242
l Non-curing silicone grease, for example, Dow Corning 4 Electrical Insulating Compound
33/57
Ranger Pro Wireless Condition Monitoring Device