User manual

SECTION 4—TYPICAL PRODUCT PARAMETERS
Perfecto
2
Series 12 Part No 1143482
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input: 3 A
Sound Level: 43 ± 2 dBA Average for Perfecto
2
Models and
Perfecto
2
V Model
40 ± 2 dBA Average for Perfecto
2
W Models
Altitude: All Perfecto
2
Models: Up to 8,000 ft (2438
meters) above sea level without degredation of
concentration levels.
All Perfecto
2
Models while filling a
HomeFill System: Up to 6,000 ft (1828
meters) above sea level without degradation of
concentration levels.
*Oxygen Output
Concentration Levels:
87% to 95.6% for flows ranging at 0.5 to 5 L/min
Maximum Outlet Pressure: 5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Flow Range: 0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Potential Obstruction Alert: 0 L/min to 0.5 L/min
The concentrator detects a condition that may
indicate a potential obstruction of the output
oxygen.
Rapid audible beeping alert (this alert is
deactivated when accessories are connected).
May be associated with flow setting of 0.5
L/min or less.
Average Power
Consumption:
Perfecto
2
and Perfecto
2
W Models: 280 W
(Operating @ 3 L/min)
Perfecto
2
V Model: 325 W
Pressure Relief Mechanism
Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum
recommended flow :
0.7 L/min (Back Pressure of 7kPa applied)
Filters: Cabinet, Outlet HEPA and Compressor Inlet