User manual

SECTION 5—OPERATING INSTRUCTIONS
Perfecto
2
Series 18 Part No 1143482
Flowrate
NOTE: For this procedure, refer to FIGURE 5.5 on page 19.
1. Turn the flowrate knob to the setting prescribed by your
physician or therapist.
CAUTION
Never set the flow greater than 5 L/min.
WARNING
DO NOT change the L/min setting on the flowmeter unless a
change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate
line on the flowmeter. Next, turn the flow knob until the ball rises to
the line. Now, center the ball on the L/min line prescribed.
NOTE: A Potential Obstruction Alert indicates a condition that
may be associated with a partial or complete obstruction of oxygen
output.
2. If the flowrate on the flowmeter ever falls below 0.5 L/min
more than about one minute, the Potential Obstruction
Alert will be triggered. This is a rapid beeping of the
audible alarm. Check your tubing or accessories for
blocked or kinked tubing or a defective humidifier bottle.
After rated flow is restored between 0.5 L/min and 0.75
L/min, the Potential Obstruction Alert will turn off.
NOTE: The use of some accessories such as the PreciseRx™
pediatric flowmeter and the HomeFill compressor will deactivate the
Potential Obstruction Alert.