PATIENT LIFT WILL NOT OPERATE WHILE CHARGING CHARGING
THE BATTERY
Ensure the power is switched “ON” (the red “RESET” button should be up).
Insert charging plug into charging port on the control box.
Plug charger to power supply.
All lights of battery indicator should be “ON” while charging.
It takes approximately 2-3 hours to fully charge the batteries from one green light.
It takes approximately 7-9 hours to fully charge the batteries from the red light.
Do not charge the batteries for an extensive period of time. This will shorten the life
of the batteries.
Do not let the batteries run down to the last red light. This might shorten the life of
the batteries or damage the batteries.
Unplug the charger first before using the lift.
KNOW WHAT YOUR BATTERY IS TELLING YOU
RED
AMBER
GREEN
GREEN
GREEN
ON
ON
ON
ON
ON
FULLY CHARGED
ON
ON
ON
OFF
OFF
NEED CHARGE
ON
ON /O FF
OFF
OFF
OFF
CHARGE IMMEDIATELY. DON'T USE THE LIFT
OFF
OFF
OFF
OFF
OFF
BAD BATTERIES OR BAD CONNECTION
WARNING BUZZERS
The control box has a buzzer that provides you with information on the
status of the batteries and the lift. To stop the buzzer, push down the RESET
button or press the Battery Test button.
BUZZER TYPE
INDICATION
NOTE
2 beeps/sec
Low Battery
(red/amber lights on)
Charge batteries immediately.
3 beeps/sec
Low Battery
(red light on)
Charge batteries immediately.
6 beeps/sec
Very Low Battery (no
light on)
Charge batteries immediately.
3 beeps every
other second
Bad Battery
(no light on)
Replace batteries.
3 long beeps
Overloaded. Batteries are charged but lift
does not lift.
Exceeded maximum load capacity.
IMPORTANT NOTES ON CHARGING