classification on the unit indicates the higher range (high-
normal in the example described).
6. Saving, retrieving and deleting
results
The results of each successful measurement are stored
together with date and time. With more than 60 items of
measured data, the earliest items of data measured are
lost.
Select the desired user memory by pressing the MEM
button and then the “+” button. Press the MEM button
again to display the average of all saved measurements in
the user memory. Press the MEM button again to display
the average day measurement for the last 7 days. (day:
5.00 a.m. – 9.00 a.m., display A). Press the MEM button
again to display the average night measurement for the
last 7 days (night: 5.00 p.m. – 9.00 p.m., display P). If you
continue pressing the MEM button, the latest individual
results are displayed with date and time.
To clear the memory, press the MEM button, then the
display shows no. 1. Press the “+” button to select the
user memory and confirm by pressing MEM again. Press
and hold the “+” and keys simultaneously for 5 sec-
onds (the display will show CLA).
If you wish to change the user memory, see “Select
memory”.
7. Cleaning and storing the instrument
Clean your device and cuff carefully only with a slightly
moistened cloth.
Do not use detergents or solvents.
On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
When storing the device, make sure that no heavy ob-
jects are placed on top of it. Remove the batteries. The
cuff tube should not have any sharp kinks
8. Error messages / troubleshooting
I
n case of faults, the E_ message appears in the display.
E
rror messages can occur when
1
. inflation pressure is higher than 300 mmHg, E2
2
. blood pressure is unusually high or low E3,
3
. you move or talk during the measurement (in addition
to E_, the cardiac arrhythmia icon appears in the
display E3),
4
. the cuff tube is not inserted properly E1,
5
. the cuff is inflated for longer than 25 seconds E1.
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk. Re-insert the batteries if necessary,
or else replace them.