Datasheet

BET-0600M
60 Watt
Desktop adapter for medical devices
Class 2 equipment – without PE
Compliant with ErP and CEC level V
Input voltage 90…264 V AC
Input frequency 50…60 Hz
Inrush current 80 A max (230 V AC)
Efficiency ≥87 %
Idle consumption ≤0.3 W (230 VAC)
Hold up time >16 ms / 115 VAC
Protection Short circuit protection, overvoltage protection, overload protection:
Switch off with auto recovery
Insulation voltage Input / Output: 6964 VDC (2x MOPP)
Line regulation ±≤2 %
Load regulation ±≤4 %
Leakage current <100 µA touch current (264 V AC / 60 Hz)
Safety / EMC TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1 3rd Edition),
CAN/CSA-C22.2 No 60601-1 3rd Edition, CE, designed according to CCC
Temperature Operating: -20…+60 °C / Storage: -20…+85 °C
Derating In the range of +40…+60 °C, 2.5 % / °C
Max. operation altitude 5000 m
MTBF > 863 000 h according to Telcordia SR-332 at +25 °C
Humidity Operating: 10…85 % RH, non-condensing / Storage: 10…90 % RH, non-condensing
Dimensions (W x D x H) 60 x 119 x 36 mm ±0.5 mm
Weight (net) 0.31 kg
Product specific data
ECO standard ErP directive 2009/125/EG, CEC level V
The series BET-0600M desktop adapters supply power to medical
devices. They are internationally certified according to IEC/
EN/UL 60601-1 standards. Due to the robust construction and
the high quality components like Japanese long life capaci-
tors, the desktop adapter series is ideally suited for many ye-
ars of reliable and safe operation. The entire closed plastic case
can easily be desinfected. Other models with different output
voltages are available on request.
0%
25%
50%
75%
100%
Ambient Temperature (˚C)
20LFM (Natural Convection)
-20 0 10 20 30 40 50 60
Derating
Cling pad, Art. No. KP-01
Medical applications
Desktop adapter
Te c h n i c a l d a t a
Bicker Elektronik GmbH | 86609 Donauwörth | Germany
Tel. +49 906 70595-0 | info@bicker.de | www.bicker.de
Extended
Extended
Fanless
Fanless
Eco
Eco
Medical
Medical
Delivery
5
YEARS
Delivery
5
YEARS
NEW
Operation
Operation

Summary of content (2 pages)