Instructions
Table Of Contents
- Index
- A Brief specification
- B Intended use
- C Technical Data
- D Name / Address / Support E-Mail / Phone number of the manufacturer
- E General Data
- E1 Assembly and installation advice
- E2 Convection and installation position
- E3 Description of connectors
- E4 Dimensioning the upstream power supply
- E5 Initial operation
- E6 Connection
- E7 Overview connector / counterpart with description / Scope of delivery
- E8 Charging time
- E9 Reverse polarity / Overload / Short circuit
- E10 Backup time
- E11 Behaviour in case of exceeding maximum backup time
- E12 Status LED
- E13 Software
- E14 Recommendations for a long UPS service life
- E15 Maintenance
- E16 Disposal
- E17 Disclaimer
- E18 Preventive measures and rules when operating the UPS system
UPSI-2406DP1 | UPSI-2406DP2
Revision 1-3
11
CONNECTION DATA INPUT / OUTPUT
Connection method Screw connection
Conductor cross-section solid 0.129 mm … 1.31 mm (26 … 16 AWG)
Conductor cross-section flexible 0.129 mm … 1.31 mm (26 … 16 AWG)
Conductor cross-section with ferrule 0.129 mm … 1.31 mm (26 … 16 AWG)
Stripping length 6 mm … 7 mm
Tightening torque 0.3 Nm … 0.4 Nm
ENERGY STORAGE UPSI-2406DP1 (BP-LIO-1525)
Charging characteristic CC / CV / CP
Nominal voltage U
N
14.8 V DC
End-of-charge voltage 15.9 V DC
Charging current 1 C max.
Deep discharge protection 10.8 V DC
Battery technology LIO
Nominal capacity 2.5 Ah (37 Wh)
Charging time 80 min
Buffer time (I
N
) 15 min
ENERGY STORAGE UPSI-2406DP2 (BP-SUC-0835)
Charging characteristic CC / CV / CP
Nominal voltage U
N
8.3 V DC
End-of-charge voltage 8.3 V DC
Charging current 15.5 A max.
Deep discharge protection 4.5 V DC
Battery technology EDLC
Nominal capacity 160 F (useful 0.972 Wh or 3.5 kJ)
Charging time 2 min
Buffer time (I
N