Specifications

8
LAMP
INSTALLATION OR
REPLACEMENT
Figure 3 - Lamp Replacement
1. Disconnect power to water purifier, and remove ground connector from the chamber.
2. Remove easy-off end cap by pulling cap off static gland nut.
3. Carefully withdraw lamp approximately 2 inches from chamber. While holding lamp end, remove lamp socket
from exposed lamp end.
4. Carefully withdraw lamp from chamber. Be sure to withdraw lamp straight out without angling until completely
clear of quartz sleeve.
CAUTION: Lamp and quartz sleeve are easily damaged. Exercise care when removing or
replacing lamp.
5. Reinstall lamp in reverse order.
WARNING: Germicidal ultraviolet rays are harmful to eyes and skin. Do not restore power to
water purifier until lamp and easy-off end caps have been properly reinstalled.
!
!
Step 1 - Remove Ground Connector
Step 2 - Remove End-Cap
Step 3 - Remove Socket
Step 4 - Remove Lamp