User Manual

AUTO CALIBRATION This procedure searches automatically for reference points within the spectra of the
lamp for reliability and reproducibility of the wavelength. The points used for
calibration are the zero order reflection of the grating (000 nm) and the Ha spectral
line of deuterium (656 nm). Auto calibration is performed automatically when power
to the detector is switched on. To access auto calibration, from the Time Constant
parameter press the Right arrow key to move to this field. Pressing the Up arrow key
starts the procedure for calibration of the wavelength scale.
LAMP The lamp type is specified using the Up arrow to toggle between deuterium and
halogen. Changing the lamp type prompts a question on whether to reset the lamp’s
working time. Pressing any green arrow key will reset the time. The lamp number is
incremented automatically and cannot be edited.
Note: See the instructions on page 5-4 for configuring the appropriate wavelength
settings in the Setup menu after a lamp change.
When replacing a deuterium lamp, the working time should be reset, as this value is
used in the GLP report. Position the cursor on the working time value and press the
Up arrow. Pressing the Up arrow again resets the working time to zero and
increments the lamp number.
The ON/OFF field enables you to switch off the lamp without switching off the
instrument.
Figure 4-3. Lamp Specification Menu
SIGNAL You can select the signal source used for all data outputs using the Up/Down arrow
keys to toggle between the following: Absorption, Signal, and Reference channels.
The values “sig” and “ref give the measured light intensity on the signal and the
reference diode respectively. The sig and ref values range between 0 and 1. Their
absolute values are dependent on the wavelength, solvent, flow cell used, and age
of the lamp.
With the dummy flow cell in place and a wavelength selection of 240 nm, the value
of reference should be between 0.1 and 0.9. A value of <0.2 indicates that the
deuterium lamp should be replaced, especially if noise and sensitivity are
unacceptable.
SCAN OUT The output speed of scan data via the integrator output. Available values are 1, 2, 5,
10, 20, and 50 steps per second (approximately. 1–50 nm/second).
USE AS A STAND-ALONE SYSTEM SYSTEM INSTALLATION AND SETUP
4-3
Link
View
Setup
Load
Prog
Hold 0
Run 1
Time
(min)
1
(nm)
2
(nm)
Edit Prog
Clear 0
Events:
off 0
2
on /
Signal 1
[ au ]
Signal 2
[ au ]
GLP