Instruction manual

QX200 Droplet Generator Instruction Manual QX200 Droplet Generator Instruction Manual
| 3
Chapter 1 QX200 Droplet GeneratorChapter 1 QX200 Droplet Generator
Table 1.2. Additional materials required for droplet generation.
Component
R
ecommended
C
atalog #
R
eagents for probe detection
PCR supermix
d
dPCR supermix for probes
1
86-3010, 186-3026, 186-3027,
1
86-3028
D
roplet PCR supermix
1
86-3023, 186-3024, 186-3025
O
ne-Step RT-ddPCR supermix for probes
1
86-3021, 186-3022
Droplet generator oil Droplet generator oil for probes 186-3030, 186-3005
Control
d
dPCR buffer control kit for probes
1
86-3052
Reagents for EvaGreen detection
PCR supermix QX200
ddPCR
EvaGreen
®
supermix 186-4033, 186-4034, 186-4035,
186-4036
Droplet generator oil
Q
X200 droplet generator oil for EvaGreen dye
1
86-4005, 186-4006
Control QX200 buffer control kit for EvaGreen dye 186-4052
Consum
ables and other materials
Pipets
2
0 μl pipet for sample loading
R
ainin L-20
5
0 μl pipet for droplet transfer
R
ainin L-50, L8-50
8
-channel, 200 μl pipet for oil
R
ainin L8-200
Pipet tips
F
iltered
R
ainin GP-L10F, GP-L200F
96-well PCR plates twin.tec semi-skirted 96-well plate Eppendorf 951020362
Reagent trough
A
ny
Foil plate seals Pierceable foil plate seals 181-4040
Plate sealer
P
X1
PCR plate sealer 181-4000
8-cap strips
A
ny
1.3 Installation and General Operation
Connect the QX200 droplet generator to a power source using only the power cord provided. Ensure the
ground is reliably connected before plugging in the instrument
Leave 10" (5 cm) clear space behind and 5" (2.5 cm) clear to the right and left for proper ventilation
Power on the droplet generator by plugging it in. The status indicator turns solid green to indicate power is on
Open and close the instrument by pressing the button on top of the green lid
U.S. standard power cord set with grounded plug
(Type 5-15P) and C5 connector (10 A/125 V)
Power supply to 5 mm DC power jack inlet