Instruction manual

QX200 Droplet Generator Instruction Manual6 |
Chapter 2 Droplet Generation
2.2 Operation of the QX200
Droplet Generator
1. Insert the DG8 cartridge into the holder with the
notch in the cartridge at the upper left of the
holder:
a. Open the cartridge holder by pressing the
latches in the middle.
b.
S
lide the DG8 cartridge into the right half of
the holder, then drop it down.
c.
P
ress the halves of the holder together to
snap it closed.
Inserting the DG8 cartridge into the cartridge holder.
The QX200 droplet generator prepares droplets for up to eight samples at a time. Droplet generation
takes ~2 minutes for each set of eight samples (~30 minutes for a 96-well plate).
All 8 sample wells in the DG8 droplet generator cartridge must contain sample (or 1x buffer
control), and all 8 oil wells must contain droplet generator oil
Do not load sample or oil into the DG8 cartridge unless it is inserted in the holder
Press
Press to open
Slide to close
Notch
Top wells (droplets)
Middle wells
(20 μl sample)
Bottom wells (70 μl oil)
DG8 cartridge