Appendix A. Maintenance
TheTrans-Blot
®
Turbo™systemrequireslittlemaintenance.AfteryounishusingtheTrans-BlotTurbo
system,rinseeachcassette,base,andlidindeionizedwatertoremoveresidualsaltsandtopreventsalt
buildup.Air-dryoruseapapertoweltodrythecassette.
Foreachtransfercassettetherearefourconnectionpinslocatedatthebackoftheinstrument
basecavity(Figure20).Overtimeitispossiblethatthesepinshavebeenexposedtobufferorother
contaminantscausingthemtostickinthedepressedposition.Withtheunitunplugged,useisopropyl
alcoholandacottonswaborisopropylwipestocleanthesepins.Pushoneachpintoensurethatit
depressesfreely.Thepinsshouldbecheckedandcleanedmonthlyaspartofstandardequipment
maintenanceprocedures.
.
Cleananyaccidentalbufferspillsintheinstrumentbayswithadrytowel.
Thecassettesandinstrumentcasingcanbecleanedwithwaterandmilddetergent;donotuse
abrasivesororganicsolvents.
Periodicallyunplugtheinstrumentandwipetheinstrumentcasingwithamoistclothorpapertowel.
Ensurethattheelectrodecontactsinthebaysareclean.Electrodeplatesshouldbeoccasionally
cleanedwithdeionizedwateroramilddetergenttoreducesaltbuildup.Ifthecathodeplatehas
excessivesaltbuildup,7%aceticacidcanbeusedtoremovetheresidue.
Occasionallycheckthatthecoolingfanventisfreeofdebrisanddust.
Fig. 20. Connection pins inside the instrument base cavity.
Connection
pins