Manual

ing protocol as described below (steps 1–7) should be used. Always reverse the flow during
this procedure so tightly-bound substances at the top of the column are quickly removed.
During this operation do not exceed more than the following maximum flow-rates.
S-1 or Q-1 1 ml/min
S-6 or Q-6 2 ml/min
S-12 or Q-12 3 ml/min
1. Wash with 2 column volumes of deionized water. Elevated backpressures may occur
when washing with deionized water. Do not exceed 700 psi.
2. Wash with 2 column volumes of 2.0 M NaCl or KCl.
3. Wash with 1 column volume of 2.0 M NaOH followed by 3 column volumes of 0.1 M
NaCl.
4. Wash with 1 column volume of 50% acetic acid followed by 3 column volumes of 0.1 M
NaCl. Wash with 1 column volume of deionized water.
5. If lipid contamination is a problem, wash with 1 column volume of MeOH followed by
3 column volumes of 0.1 M NaCl.
6. Wash with 2 column volumes of 2.0 M NaCl or KCl or the salt containing the desired
counter-ion.
7. Change the top frit and wash with 3 column volumes of start buffer.
Fig. 1. Drawing of UNO column.
6
Tubing, .020 I.D. x 15 cm (1)
Tube end fitting, 1/4 x 28 (2)
End cap (3)
Shield (4)
Bed support (5)
O-ring (6)
Spacer (7)
Distribution screen (8)
Frit (9)
Glass tube (10)
Frit (9)
Distribution screen (8)
Spacer (7)
O-ring (6)
Bed support (5)
End cap (3)
Tube end fitting, 1/4 x 28 (2)
Tubing, .020 I.D. x 15 cm (1)
UNO Column