4.6 TRANSPORT AND STORAGE CONDITIONS
Temperature
Atmospheric
pressure
Relative humidity
Transport
–10°C to +40°C
(14°F < T° < 104°F)
500 hPa to 1060 hPa 10% to 90%
Storage
–10°C to +40°C
(14°F < T° < 104°F)
800 hPa to 1060 hPa 10% to 90%
Table 11 - Transport and storage conditions
CAUTION
- The device must be transported and stored in its original packaging designed to protect it
from damage.
- Storage and transport conditions must meet conditions described in the above table.
4.7 USAGE CONDITIONS
Temperature
Atmospheric
pressure
Relative humidity
Operating
+15°C to +35°C
(59°F < T° < 95°F)
800 hPa to 1060 hPa 30% to 70%
Table 12 - Usage conditions
CAUTION
- Leave the device in a room for 30 minutes before unpacking to ensure there is no
condensation.
4.8 DISPOSAL
Dispose of EMBODY according to your local guidelines.
According to Directives 2012/19/UE WEEE and 2011/65/UE RoHS II on the restriction of
hazardous substances in electrical and electronic equipment on their disposal:
Public authorities adopt adequate measures to make sure that users, distributors and
manufacturers contribute to the collection of electrical and electronic equipment, setting legal
requirements to reuse, recover or recycle the said equipment.
CAUTION
You must take into account the potentially harmful effects to the environment or human health
due to the improper disposal of the equipment or of parts of it.
The following graphic symbol is applied on the label of the equipment:
It reminds that all electrical and electronic equipment must be collected and disposed of
separately from domestic waste at the end of life.
CAUTION
Before disposing of EMBODY: reset the device to delete all your personal health data (see
section 8.2.)