CALIBRATING YOUR EMBODY
Disconnect the USB cable from
the EMBODY base.
Check that the EMBODY LED is
flashing orange.
Slowly tilt the EMBODY base side
to side to CALIBRATE.
Continue tilting until:
The base beeps and the LED
light turns green.
Then the light will turn off.
NOTE:
If the orange LED does not light up after disconnecting the USB cable from the EMBODY
base, it means that the calibration was done automatically. You can go directly to the next
section:
Section 5.3. Assembling your EMBODY to the insulin pen.