Page 13 / 27 IFU_EMB_AD_US-CA-EN_V2
5.3 ASSEMBLING YOUR EMBODY TO THE INSULIN PEN
CAUTION
The EMBODY base and button must be correctly attached onto the pen, otherwise, data will
not be transferred.
ATTACHING BASE TO PEN
Hold the EMBODY base so
that the clear square plastic
window is facing you
Align the EMBODY base
clear square plastic window
with the pen dose window,
then snap the base onto the
pen.
Make sure the EMBODY
base window is perfectly
aligned with the pen dose
window.
NOTE:
A pen symbol is represented on the
side of the EMBODY
base to help you
position the pen in the correct
orientation (square window oriented
toward the top of the pen).
ATTACHING BUTTON TO PEN
Align the EMBODY button’s grooves with
the dose injection button’s notches, then
insert the EMBODY button onto the pen
injection button. Make sure the button is
properly attached and fully inserted.
The EMBODY base LED will flash
orange and green when the button and
base are properly mounted onto the pen
and when the EMBODY device is not yet
paired.