Use Instructions

Page 16 / 27 IFU_EMB_AD_US-CA-EN_V2
6.2 INJECTION
WARNING
If you are not certain that you injected your insulin, do not start or repeat your injection.
Monitor your blood glucose as instructed by your healthcare provider.
IN AUTONOMOUS MODE OR
PAIRING MODE
=>
follow the steps indicated below:
Check the insulin dose to inject according to the prescription and the type of
medication.
Turn the button to select the number of units to inject. The dose indicator should
line up with the dose to inject.
Press on the injection button with your thumb until you return to ZERO (complete
dose injection)
Keep the injection button pressed in
and the needle in the skin:
- While the EMBODY’s green LED is
flashing,
- Until EMBODY emits an audible signal
(BEEP) signalling the end of the injection.
You can release the pressure on the button and then remove the needle from the
skin when the green LED stops flashing and the “BEEP” signaling the end of
injection is emitted.
WARNING
If you do not see ZERO (“0”) in the pen dose window: it indicates that the full dose has not
been injected.
IN PAIRING MODE ONLY
When the device is coupled with a compatible mobile application installed on a smartphone:
The application automatically detects an injection.
The information corresponding to the injected dose is displayed: dose, day and time.
Please refer to the application manufacturer’s instructions.
COMPATIBLE MOBILE APP. SCREENS INFORMATION
Refer to the manufacturer’s application instructions for use for more information regarding
the functionalities of the application you are using. More particularly:
- To display the selected and injected insulin dose
-
To display the date and time of each injection.