ORANGE LED – TROUBLESHOOTING RECOMMENDATIONS
The EMBODY base light indicator (LED) remains orange if:
- the clear square plastic window of the EMBODY
base and the dose window of the pen are not
correctly aligned, or
- if the base is mounted too low / too high ( the
top of the EMBODY base must be aligned with
the top of the pen, at the base of the dose button
on the pen), or
- if the EMBODY button is not pushed on enough.
In this case, it is necessary to:
- Adjust and align:
the position of the windows:
or raise or lower the position of the EMBODY base on the pen,
- Or, to push the EMBODY button onto the pen,
until the EMBODY base LED flashes green and orange if EMBODY is not paired (if EMBODY
is paired, the light indicator is green).